Frequency-comb-referenced multi-wavelength profilometry for largely stepped surfaces.

Optics Express, Vol. 21, Issue 8, pp. 9780-9791 (2013).

Sangwon Hyun, Minah Choi, Byung Jae Chun, Seungman Kim, Seung-Woo Kim, and Young-Jin Kim.

 

Ultrafast Optics for Ultraprecision Group, Department of Mechanical Engineering, Korea Advanced Institute of Science and Technology (KAIST),Science Town, Daejeon, 305-701, South Korea.

 

 

Abstract

 

3-D profiles of discontinuous surfaces patterned with high step structures are measured using four wavelengths generated by phase-locking to the frequency comb of an Er-doped fiber femtosecond laser stabilized to the Rb atomic clock. This frequency-comb-referenced method of multi-wavelength interferometry permits extending the phase non-ambiguity range by a factor of 64,500 while maintaining the sub-wavelength measurement precision of single-wavelength interferometry. Experimental results show a repeatability of 3.13 nm (one-sigma) in measuring step heights of 1800, 500, and 70 um. The proposed method is accurate enough for the standard calibration of gauge blocks and also fast to be suited for the industrial inspection of microelectronics products.

© 2013 OSA

Go To Journal

 

Frequency-comb-referenced multi-wavelength profilometry for largely stepped surfaces]