Sensors and Actuators B: Chemical, Volume 173, October 2012, Pages 157-163
Hans-Peter Loock, Peter D. Wentzell
Department of Chemistry, Queen’s University, Kingston, ON K7L 3N6, Canada
Department of Chemistry, Dalhousie University, Halifax, NS B3H 4J3, Canada
Abstract
The limit of detection (LOD) and the sensitivity of a chemical sensor are defined using IUPAC guidelines. The LOD from simulated and experimental data is calculated from a calibration curve using a simple statistical model that was implemented into a spreadsheet program. This definition of the LOD is compared with the commonly used definition of the LOD, which is based on the product of sensitivity and the theoretical instrument resolution.
Advances in Engineering Advances in Engineering features breaking research judged by Advances in Engineering advisory team to be of key importance in the Engineering field. Papers are selected from over 10,000 published each week from most peer reviewed journals.